Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.
LifeCare PCA 3 Infusion System (March 07, 2013)
- Starting date:
- March 07, 2013
- Posting date:
- March 13, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Product Safety
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-26337
Reason
The Lifecare PCA Infusion Pumps have a clear plastic door that closes and locks to deter access to the syringue/vial used to deliver the medication. The door can be flexed to the point which allows unauthorized personnel to use objects to press the vial injector or the vial to receive an un-prescribed bolus of medication.
Affected products
LifeCare PCA 3 Infusion System (2013-03-07)
Lot or serial number
All serial numbers
Model or catalog number
12384, 20709