Health product recall

Treatment Table for Elekta Precise (2014-08-21)

Starting date:
August 21, 2014
Posting date:
August 28, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41211

Recalled products   

  1. Treatment Table for Elekta Precise

Reason

Elekta has identified that it is possible to get a positional error with the precise treatment table. This can occur if the positional sensors are not correctly installed during corrective maintenance.

Affected products

A. Treatment Table for Elekta Precise

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • MRT 5071
Companies
Manufacturer
Elekta Limited
Linac House, Fleming Way
Crowley, Sussex
RH10 9RR
UNITED KINGDOM