Treatment Table for Elekta Precise (2014-08-21)
- Starting date:
- August 21, 2014
- Posting date:
- August 28, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-41211
Recalled products
- Treatment Table for Elekta Precise
Reason
Elekta has identified that it is possible to get a positional error with the precise treatment table. This can occur if the positional sensors are not correctly installed during corrective maintenance.
Affected products
A. Treatment Table for Elekta Precise
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- MRT 5071