Mammomat Inspiration Mammography System with Stereostatic/Biopsy Device (2014-09-22)

Starting date:
September 22, 2014
Posting date:
October 1, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41599

Recalled Products

Mammomat Inspiration Mammography System with Stereostatic/Biopsy Device

Reason

The stereo biopsy devices (serial numbers 2000 through 2090) for Mammomat Inspiration Mammography Systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch being set sideward. If this occurs with the needle being already inserted, it could lead to an unintentional patient injury.

Affected products

Mammomat Inspiration Mammography System with Stereostatic/Biopsy Device

Lot or serial number

10140000-1146

10140000-2042

Model or catalog number