Health product recall

Coulter HMX Hematology Analyzers and Coulter IH 500 Series System (2014-10-20)

Starting date:
October 20, 2014
Posting date:
October 29, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41977

Recalled Products

A) Coulter HMX Hematology Analyzers
B) Coulter IH 500 Series System

Reason

Customers have observed that the differential results for coulter 5c abnormal ii cell control recovered an excessive number of incomplete computation (.....) And/or review ("R") flags on the coulter IH 500 series and HMX analyzers. The issue may occur more often when laboratory temperatures fall below 70 degrees Fahrenheit (21 degrees Celsius).

Affected products

A) Coulter HMX Hematology Analyzers

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

6605523
6605526

Companies
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES

B) Coulter IH 500 Series System

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

178833
 

Companies
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES