Health product recall

Quanta Flash DGP IGA calibrators (2015-03-03)

Starting date:
March 3, 2015
Posting date:
March 27, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-52701

Recalled Products

A. Quanta Flash DGP IGA calibrators

Reason

The kit box and calibrator tubes are correctly labelled with Lot 141010 but the display and reports on the bio-flash system indicate Lot "141009". This was caused by an error in the barcoding of the calibrator tubes, which also assigned target values to the calibrators from lot 141009 rather than from Lot 141010.  The target values of the two Lots are comparable. Therefore, results previously reported are not impacted by this recall.

Affected products

A. Quanta Flash DGP IGA calibrators

Lot or serial number

141010

Model or catalog number

701166

Companies
Manufacturer
Inova Diagnostics Inc.
9900 Old Grove Road
San Diego
92131
California
UNITED STATES