This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

ARTIS DIALYSIS SYSTEM(2015-09-17)

Starting date:
September 17, 2015
Posting date:
October 8, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55296

Affected Products

SYSTÈME DE DIALYSE ARTIS

Reason

There have been reports of a burning smell/smoke coming out from the upper ventilation grids of the Artis dialysis machine. The events have been associated with a limited number of power supply units manufactured with a specific batch of capacitor. The technical analysis identified that the issue was caused by a defective capacitor inside the power supply unit which has the potential to melt after approximatively 4 years of use.

Affected products

ARTIS DIALYSIS SYSTEM

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

112949

Companies
Manufacturer
Gambro Dasco S.P.A.
Via Modenese 66
Medolla
ITALY