P-300 PORTABLE CEILING LIFT AND C-300 FIXED CEILING LIFT (2015-10-20)
- Starting date:
- October 20, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-55772
Affected Products
- P-300 PORTABLE CEILING LIFT
- C-300 FIXED CEILING LIFT
Reason
Prism Medical Canada has been made aware of a number of incidents globally involving 300 Series Lifts in which the drive gear shaft has broken, eliminating the lifts’ ability to raise or lower a client. Of these incidents they have discovered very few cases in which the free fall protection system did not engage properly. If both of these incidents occur at the same time there is a risk for the person being supported by the lift to drop unexpectedly.
Affected products
A. P-300 PORTABLE CEILING LIFT
Lot or serial number
All lots
Model or catalog number
- 303050
- 303050R
- 303051
Companies
- Manufacturer
-
Prism Medical USA
45 Progress Parkway
Maryland Heights
63043
Maryland
UNITED STATES
B. C-300 FIXED CEILING LIFT
Lot or serial number
All lots
Model or catalog number
- 330050
- 330050R
- 330053
- 330053R
- 330057
Companies
- Manufacturer
-
Prism Medical USA
45 Progress Parkway
Maryland Heights
63043
Maryland
UNITED STATES