SYNGO DYNAMICS VA10 VERSION (2016-02-01)
- Starting date:
- February 1, 2016
- Posting date:
- February 12, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-57114
Affected Products
SYNGO DYNAMICS VA10 VERSION
Reason
Siemens is notifying syngo Dynamics customers of a potential issue when using the measurement package of the VA10 version. If two measurements with the same name exist, instances of measurement will populate both measurements definitions. One of the definitions can be edited while one cannot be seen or edited. Both set of measurements are used in any associated calculations. Therefore if you edit or delete one of the measurement instances the one that is hidden will not receive the edits and all instances, edited and not edited, will be averaged together.
Affected products
SYNGO DYNAMICS VA10 VERSION
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
VA10
Companies
- Manufacturer
-
Siemens Healthcare GMBH
HENKESTRASSE 127
ERLANGEN
91052
GERMANY