This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Offset Reamer Handle

Starting date:
November 14, 2016
Posting date:
November 25, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-61224

Reason

Greatbatch Medical identified an issue with its Offset Reamer Handle, as field complaints were reported that the drive chain mechanism seizes during use. The function of the drive chain mechanism is to transmit rotational power from a mechanical drill to an acetabular reamer during total hip arthroplasty. During seizing, the device is rendered non-functional in the operating room.

Affected products

Offset Reamer Handle

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

00780408000

Companies

Manufacturer
Greatbatch Medical
10000 Wehrle Drive
Clarence
14031
UNITED STATES