Plum 360 Infusion Pump Software
- Starting date:
- January 4, 2017
- Posting date:
- January 18, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-61848
Reason
The pump software incorrectly generates a depleted battery alarm instead of a replace battery alarm when the battery has a near end of life charge capacity.
Affected products
Plum 360 Infusion Pump Software
Lot or serial number
Software Version 15.02
Model or catalog number
30010
Companies
- Manufacturer
-
Hospira Inc.
275 North Field Drive
Lake Forest
60045
Illinois
UNITED STATES