SIGMA SPECTRUM INFUSION PUMP W/MASTER DRUG LIBRARY (V6) (2017-10-04)

Starting date:
October 4, 2017
Posting date:
October 11, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-64818

Affected products

SIGMA SPECTRUM INFUSION PUMP W/MASTER DRUG LIBRARY (V6)

Reason

Baxter Corporation is issuing an urgent medical device correction for certain Sigma Spectrum Infusion pumps with Master Drug Library (v6) due to the potential that a small number of these pumps may exhibit excessive wear of the pumping mechanism caused by a lack of lubrication. This could potentially result in over- or under-infusion.

Affected products

SIGMA SPECTRUM INFUSION PUMP W/MASTER DRUG LIBRARY (V6)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

35700BAX