Health product recall

Optilite Beta-2-microglobulin Urine Kit (2018-03-09)

Starting date:
March 9, 2018
Posting date:
April 3, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66328

Affected products

OPTILITE BETA-2 MICROGLOBULIN URINE KIT

Reason

Following an external complaint, investigations were reviewed on the 26/02/2018 and concluded the antigen excess protection of the assay at the initial dilution is only up to 18.4 mg/L (doc. 1 IFU). This does not meet the antigen excess protection (760 mg/L) stated in the IFU.

Affected products

OPTILITE BETA-2 MICROGLOBULIN URINE KIT

Lot or serial number

407459

Model or catalog number

LK043.L.OPT

Companies
Manufacturer

The Binding Site Group Limited

8 Calthorpe Road

Birmingham

B15 1QT

UNITED KINGDOM