Health product recall

Universal Shears 7 1/2" (2018-06-13)

Starting date:
June 13, 2018
Posting date:
June 29, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67126

Affected products

Universal Shears 7 1/2"

Reason

Teleflex medical has issued a recall for two lots of universal shears (product code 142452 lots F7 and G7). The incorrect plastic was used to manufacture the handles of these shears, which causes the handles to melt if sterilized.

Affected products

Universal Shears 7 1/2"

Lot or serial number
  • F7
  • G7
Model or catalog number
  • 142452
Companies
Manufacturer
Teleflex Medical
2917 Weck Drive
Research Triangle Park
27709
UNITED STATES