Health product recall

Gentec Medical Air Regulators - 287 MA Series (2018-06-27)

Starting date:
June 27, 2018
Posting date:
July 13, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67222

Affected products

Gentec Medical Air Regulators - 287 MA Series

Reason

At the outlet of the Regulator, a DISS Oxygen fitting might be mistakenly installed instead of a DISS Air fitting.  Users will not be able to connect Air hose to the outlet.  In addition, there is wording "OXYGEN" printed on gauge face. This may mislead users that the regulators supply Oxygen gas.

Affected products

Gentec Medical Air Regulators - 287 MA Series

Lot or serial number

All lots.

Model or catalog number

GEN-287MA-25LY1

GEN-287MA-25LY2

Companies
Manufacturer

GENSTAR TECHNOLOGIES CO., INC.

4525 Edison Ave

Chino

91710

California

UNITED STATES