Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.

Oasis Drain, Dry Single W/AC (2015-05-07)

Starting date:
May 7, 2015
Posting date:
June 1, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53577

Recalled Products

  1. Oasis Drain, Dry Single W/AC

Reason

In rare cases the O-ring on the in-line connector may be missing on the end of the patient line. The O-ring is a black circular ring that is clearly visible when the two pieces of the in-line connector are separated.  If this O-ring is missing, then air can pass into the patient tubing causing a system air leak.  The chest drain system will no longer be a closed system as required and you will not obtain appropriate function of the drain in evacuating air, blood or fluid.

Affected products

A. Oasis Drain, Dry Single W/AC

Lot or serial number
  • L/N 222115
  • L/N 222118
  • L/N 222528
Model or catalog number

3600-100

Companies
Manufacturer
Atrium Medical Corporation
5 Wentworth Drive
Hudson
New Hampshire
UNITED STATES