Health product recall

Revolution CT (2018-10-17)

Starting date:
October 17, 2018
Posting date:
November 15, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68218

Affected products

Revolution CT

Reason

GE Healthcare has become aware of a localization error on revolution CT systems equipped with the smartstep option. The z location displayed on image annotation and on the gantry display may not be the actual location of the table. This can occur if:

  1. The table has been moved all the way out of the gantry using the 'out' button on the gantry control or the 'unload pedal'.
  2. The table is then moved back into the gantry using the 'move to start location' button on the hand held controller.

In this situation, if a scan is initiated by the user, the scan may be acquired in the wrong location. An additional scan may be necessary.

Affected products

Revolution CT

Lot or serial number
  • 00000440051CN1
  • 00000441839CN8
  • 00000442219CN2
  • 00000443342CN1
  • REVVX1600054CN
  • REVVX1700084CN
Model or catalog number
  • 5590000
  • 5590000-6