Health product recall

Alinity i Processing Module (2018-12-26)

Starting date:
December 26, 2018
Posting date:
January 23, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68844

Affected products

Alinity i Processing Module

Reason

Abbott has identified an issue with the Alinity I gear pump assembly (part number a 30108552-01) resulting in foaming/bubbling out of the bottle reservoir for concentrated wash buffer and an unexpected amount of dried residue of buffer. The foaming/bubbling is a result of an improper calibration of the pump's operating speed.

Affected products

Alinity i Processing Module

Lot or serial number
  • Ai01457
  • Ai01459
  • Ai01001 to Ai01455
Model or catalog number

03R65-01

Companies
Manufacturer

Abbott GmbH and Co. K.G.

Max Planck Ring 2

Wiesbaden

65205

GERMANY