Health product recall

enFlow Disposable Cartridge (2019-03-14)

Starting date:
March 14, 2019
Posting date:
March 21, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69350

Affected Products

enFlow Disposable Cartridge

Reason

Vyaire's decision to initiate this urgent recall notification was based on recently conducted internal testing which has indicated that there is the potential to elute aluminum from the enFlow Disposable Cartridge during intravenous warming therapy with fluid and blood solutions.  This global recall is being conducted based on the potetial safety risk associated with aluminium toxicity.

Affected products

enFlow Disposable Cartridge

Lot or serial number

All lots.

Model or catalog number
  • 980200EU
  • 980202EU
Companies
Manufacturer
Vyaire Medical (Formerly Trading as CareFusion)
22745 Savi Ranch Parkway
Yorba Linda
92887
California
UNITED STATES