Health product recall

HeartStart MRx defribrillator/monitor (2019-05-27)

Starting date:
May 27, 2019
Posting date:
June 7, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70107



Last updated: 2019-06-07

Affected Products

HeartStart MRx defribrillator/monitor

Reason

If the HeartStart MRx is used without a charged M3538A lithium ion battery installed, the device may not operate if AC mains power is interrupted or if the AC Power Module does not function properly. If the M3539A ac power module fails, or AC mains power to the HeartStart MRx is otherwise interrupted, without a charged M3538A lithium ion battery installed, as directed in the device's instructions for use, the HeartStart MRx may lose all power and fail to operate. A faulty AC Power Module may also fail to charge the battery.

Affected products

HeartStart MRx defribrillator/monitor

Lot or serial number

All lots.

Model or catalog number

M3535A

M3536A

Companies
Manufacturer

Philips Medical Systems

3000 Minuteman Road

Andover

01810

Massachusetts

UNITED STATES