Symbia T series Gamma Camera/CT Scanners (2019-07-12)
- Starting date:
- July 12, 2019
- Posting date:
- July 26, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70527
Last updated: 2019-07-26
Affected Products
Symbia T series Gamma Camera/CT Scanners
Reason
A defect in the broad quantification option of the Symbia product software version vb20a may cause the system not to retain:
- The phantom volume entered by the user during volume sensitivity calibration.
- The residual dose entered by the user during volume sensitivity calibration or acquisition.
The failure only occurs when broad quantification is purchased without other xSPECT Quant options. The software update vb20b service pack will eliminate the issue, correct other minor software problems and improve performance.
Affected products
Symbia T series Gamma Camera/CT Scanners
Lot or serial number
Not applicable.
Model or catalog number
10275007
10275008
10275009
Companies
- Manufacturer
-
Siemens Medical Solutions USA, Inc. Molecular Imaging
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES