Health product recall

Giraffe Incubator, Giraffe Omnibed, Giraffe Incubator Carestation, and Giraffe Omnibed Carestation (2019-10-22)

Starting date:
October 22, 2019
Posting date:
November 1, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71481



Last updated: 2019-11-01

Affected Products

  1. Giraffe OmniBed
  2. Giraffe Incubato
  3. Giraffe OmniBed Carestation
  4. Giraffe Incubator Carestation,

Reason

The bedside panels of Giraffe Incubator, Giraffe Omnibed, Giraffe Incubator Carestation, and Giraffe Omnibed Carestation can be upright and look closed but not be latched.  The portholes also can look closed when not latched. If a canopy cover is used, it can hold the bedside panel or porthole door closed without being latched. If a bedside panel or porthole that is not latched falls open, it will no longer protect the patient from falling.

Affected products

A. Giraffe OmniBed

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

2074920-001-XXXXXX

OMNIBED

Companies
Manufacturer

Datex-Ohmeda, Inc

9900 Innovation Drive

Wauwatosa

53226

Wisconsin

UNITED STATES


B. Giraffe Incubator

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number

2074921-001-XXXXXX

GIRAFFE INCUBATOR

Companies
Manufacturer

Datex-Ohmeda, Inc

9900 Innovation Drive

Wauwatosa

53226

Wisconsin

UNITED STATES


C. Giraffe OmniBed Carestation

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

2082844-001

Companies
Manufacturer

Datex-Ohmeda, Inc

9900 Innovation Drive

Wauwatosa

53226

Wisconsin

UNITED STATES


D. Giraffe Incubator Carestation,

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

2082844-002

Companies
Manufacturer

Datex-Ohmeda, Inc

9900 Innovation Drive

Wauwatosa

53226

Wisconsin

UNITED STATES