Artis Pheno System (2020-05-23)

Starting date:
May 23, 2020
Posting date:
May 23, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73149

Last updated:

Affected Products

Artis Pheno System

Reason

In the event of a fault within the drivetrain of an axis, the potential exists for the C-arm to leave its intended travel path by up to 10cm. The problem has not been reported at customer sites and is currently classified as a single event.

Affected products

Artis Pheno System

Lot or serial number

164101
164123
164221

Model or catalog number

Not applicable.

Companies
Manufacturer
Siemens Healthcare GMBH
Henkestr. 127
Erlangen
91052
GERMANY