epoc BGEM Test Card (2020-09-02)
- Starting date:
- September 2, 2020
- Posting date:
- September 18, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73927
Last updated: 2020-09-18
Affected Products
epoc BGEM Test Card
Reason
Siemens Healthcare Diagnostics has identified the occurrence of sporadically inconsistent discrepant (low bias) glucose results on card lot 01-20095-10. For this lot, an occurrence rate of up to 17.5% discrepant (low bias) glucose results is possible with maximum observed biases.
Affected products
epoc BGEM Test Card
Lot or serial number
01-20095-10
Model or catalog number
10736515
Companies
- Manufacturer
-
Epocal Inc.
2060 Walkley Road
Ottawa
K1G 3P5
Ontario
CANADA