HLS Set Advanced 5.0/7.0 with BIOLINE Coating (2020-11-10)
- Starting date:
- November 10, 2020
- Posting date:
- November 20, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-74359
Last updated:
Affected Products
HLS Set Advanced 5.0/7.0 with BIOLINE Coating
Reason
During verification testing and a review of customer complaints for the HLS Set Maquet Cardiopulmonary (MCP) identified a potential impairment of the sterile packaging barrier. Under the transport simulation verification tests, movements of the device and its accessories inside the plastic tray have been identified. As a result the sterile barrier for the system may be compromised.
Affected products
HLS Set Advanced 5.0/7.0 with BIOLINE Coating
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
BEQ-HLS 5050-CA
BEQ-HLS 7050-CA
Companies
- Manufacturer
-
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY
- Date modified: