Oxford Partial Knee System Fixed Lateral Tibial Construct, Right-Cemented, E3 (2020-12-02)
- Starting date:
- December 2, 2020
- Posting date:
- December 18, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-74567
Last updated: 2020-12-18
Affected Products
Oxford Partial Knee System Fixed Lateral Tibial Construct, Right-Cemented, E3
Reason
Zimmer Biomet is conducting a lot specific medical device recall for multiple sterile Zimmer Biomet products due to the potential for a breach in the sterile seal. A breach in the packaging seal could potentially lead to loss of sterility of the device. Based on an inspection of the affected packaging, a breach in the sterile seal is estimated to occur in less than 0.5% of the affected product.
Affected products
Oxford Partial Knee System Fixed Lateral Tibial Construct, Right-Cemented, E3
Lot or serial number
207230
Model or catalog number
154370