LifePak CR2 Defibrillator (2021-02-03)
- Starting date:
- February 3, 2021
- Posting date:
- February 9, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type I
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-74973
Last updated: 2021-02-09
Affected Products
LifePak CR2 Defibrillator
Reason
Stryker has become aware that the lid magnet for this device can occasionally become separated and lost. Until recently, Stryker was previously only aware that if the lid magnet is missing, the device would not automatically turn on and off when the lid is opened and closed. Stryker has now become aware that if the lid magnet is missing, it also causes the LIFEPAK CR2 battery to deplete more quickly resulting in premature battery life. (Complaint trending from April 30 2020 to November 1 2020). The LPCR2 Operating Instructions currently provide users with no warning that the battery can become prematurely depleted if the lid magnet is missing.
Affected products
LifePak CR2 Defibrillator
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
More than 10 numbers, contact manufacturer.
Companies
- Manufacturer
-
Physio-Control, Inc.
11811 Willows Road North East
Redmond
98052
Washington
UNITED STATES