Health product recall

Leckey MyWay Gait Trainer/Walker (2021-03-08)

Starting date:
March 8, 2021
Posting date:
March 25, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75199

Last updated: 2021-03-25

Affected Products 

Leckey MyWay Gait Trainer/Walker

Reason

The rear legs of the MyWay walker each have a boss into which castors are screwed. This boss can become loose causing the castor to loosen from the rear leg of the chassis. Investigations have confirmed that this is due to variability in the weld process used to attach the boss to tube.  The manufacturer has identified a potential for patient harm.  In the interest of user safety, a decision has been taken to issue a Field Safety Notice proactively (FSN-01-03-21). This will involve a MyWay Walker Rear Leg replacement to be carried out in the field.

Affected products

Leckey MyWay Gait Trainer/Walker (2021-03-08)

Lot or serial number

LEC/262983
LEC/264958
LEC/267877

Model or catalog number

173-1600
173-2600
173-3600

Companies
Manufacturer
JAMES LECKEY DESIGNS LTD
A BALLINDERRY ROAD,
LISBURN
BT2825A
IRELAND