GUNTHER TULIP RETRIEVAL SET (2021-04-08)

Starting date:
April 8, 2021
Posting date:
April 22, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75439

Last updated:

Affected Products 

GUNTHER TULIP RETRIEVAL SET

Reason

Cook Inc. has identified that these lots may contain an 0.038" wire guide instead of the specified 0.035" wire guide, resulting in an incompatibility with the supplied dilator (dilator cannot be advanced over the wire guide).

Affected products

GUNTHER TULIP RETRIEVAL SET

Lot or serial number

13477341

Model or catalog number

GTRS-200-RB

Companies
Manufacturer
COOK, INC
750 DANIELS WAY
BLOOMINGTON
47404
Indiana