PINNACLE ACETABULAR CUP SYSTEM (2021-05-13)
- Starting date:
- May 13, 2021
- Posting date:
- May 26, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- Healthcare Professionals, General Public, Hospitals
- Identification number:
- RA-75665
Last updated: 2021-05-26
Affected Products
PINNACLE ACETABULAR CUP SYSTEM
Reason
There is a potential of an out of specification condition (oversized "minor diameter") on certain PINNACLE® CUP devices. This correction is intended to advise physicians, who use the Apex HE, of the oversize condition of the hole in the cup, which could lead the Apex HE to thread through the shell of the cup without stopping or to protrude internally as a result of "cross-threading". This may occur if the Apex HE is manipulated or toggled while threading it through the hole of certain PINNACLE® CUP.
Affected products
PINNACLE ACETABULAR CUP SYSTEM
Lot or serial number
- 9675033
Model or catalog number
- 1217-32-050
Companies
- Manufacturer
-
Depuy Orthopaedics Inc.
700 Orthopaedics Drive
Warsaw
46582
Indiana
UNITED STATES