Spectrum IQ Infusion System with Dose IQ Safety Software (2021-06-10)
- Starting date:
- June 10, 2021
- Posting date:
- June 23, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-75873
Last updated: 2021-06-23
Affected Products
Spectrum IQ Infusion System with Dose IQ Safety Software
Reason
Baxter has received customer reports of watchdog system errors which may occur following frequent connection and disconnection from the facility’s wireless network. This leads to elevated stress on the pump’s memory and processing functions, which causes instability and triggers the system error. The alarm causes the pumps to enter a fail-safe mode which stops all pump processes including delivery of fluid to the patient.
Affected products
Spectrum IQ Infusion System with Dose IQ Safety Software
Lot or serial number
All serial numbers.
Model or catalog number
3570009
Companies
- Manufacturer
-
Baxter Healthcare Corporation
1 Baxter Parkway
Deerfield
60015
Illinois
UNITED STATES