Affected Products
Data-Cyte Plus
Reason
During the manufacturer routine vigilance process, it has been detected that cell 10, (donor M6050AM) diluted to 0.8% for data-cyte plus 0.8%, ref. 213587 Lot 610021015 exp. 2021-09-04, Despite being typed as FYA+ (positive) and FYB+ (positive) during the bulk batch release and confirmed as positive in the final product release testing, has lost its reactivity over the time for these two antigens.