ALINITY M SARS AMP KIT (2021-09-07)
- Starting date:
- September 7, 2021
- Posting date:
- September 17, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-76443
Last updated: 2021-10-15
Affected Products
A. ALINITY M SARS-COV-2 AMP KIT
B. ALINITY M RESP-4 PLEX AMP KIT
Reason
Abbott Molecular inc. Has received reports of Alinity M SARS-CoV-2 EUA false positive results. Preliminary investigation has identified that potential carryover in the assay tray may contribute to false positive results. An evaluation of the current mixing protocol used during preparation of the PCR reaction mixture determined the current mixing parameters may result in potential overflow that could carry over into neighboring wells in the assay reagent tray. Because the assay parameters and the presence of the SARS-CoV-2 analyte are similar, Alinity M Resp-4-Plex customers are being included in this communication.
Affected products
ALINITY M SARS-COV-2 AMP KIT
Lot or serial number
All lots
Model or catalog number
09N78-090
Companies
- Manufacturer
-
Abbott Molecular Inc.
1300 E. Touhy Ave.
Des Plaines
60018
Illinois
UNITED STATES
ALINITY M RESP-4 PLEX AMP KIT
Lot or serial number
All lots
Model or catalog number
09N79-090
Companies
- Manufacturer
-
Abbott Molecular Inc.
1300 E. Touhy Ave.
Des Plaines
60018
Illinois
UNITED STATES