ORT (2021-09-20)
- Starting date:
- September 20, 2021
- Posting date:
- October 1, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-76559
Last updated: 2021-10-01
Affected Products
ORT300
Reason
The ORT200/300 operating room table is specifically designed to provide an mr compatible platform for neurosurgical use for cranial and spinal surgeries. The tables are removable from the operating room and mate to a docking station installed in the floor of the or via a floor latch mechanism which contains 4 linear actuators. Complaints have reported an inability to dock or undock the operating room table from the floor baseplate, due to malfunction of the actuators within the floor latch locking mechanism.
Affected products
ORT300
Lot or serial number
Not applicable.
Model or catalog number
114093-600
Companies
- Manufacturer
-
Deerfield Imaging, Inc.
5101 Shady Oak Road
Minnetonka
55343
Minnesota
UNITED STATES