Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.

M11 Ultraclave Steam Sterilizer (March 15, 2012)

Starting date:
March 15, 2012
Posting date:
April 30, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Product Safety
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72543

Reason

The recall has been issued due to the Sterilizer's door inadvertently opening, while the unit is cycling. This occurs when the user does not properly latch the door and the unit fails to detect the door as improperly closed. The failure of the detection mechanism has been attributed to one of the following failures. Door switch failed closed, short in the capacitor across the door switch, short across switch leads, firmware error or intentional tampering.

Affected products

M11 Ultraclave Steam Sterilizer(2012-03-15)

Lot or serial number

All lots

Model or catalog number

M11-001, M11-001R, M11-002, M11-002R, M11-003, M11-004, M11-005, M11-006, M11D-001, M11D-002