Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada, you can request alternate formats by contacting us.
OneTouch Verio IQ Blood Glucose Meter (March 25, 2013)
- Starting date:
- March 25, 2013
- Posting date:
- March 25, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Product Safety
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-26407
Reason
The Onetouch Verio IQ will not work as expected at extremely high blood glucose levels of 56.8 Mmol/L (1023 mg/dL) or above. The Onetouch Verio IQ will not display a result and will not write to memory. The meter will countdown then turn off instead of displaying a result.
Affected products
OneTouch Verio IQ Blood Glucose Meter (2013-03-25)
Lot or serial number
All lots
Model or catalog number
011-838, 022-276 (System Kit), 022-277 (Starter Kit), 022-282 (Warranty)